8. Operating Instructions – PRO Card and Modem
8-14 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
8.4 Install the Modem Option
In addition to the PRO Card, your HomeChoice PRO APD System can also transfer
therapy information via a phone line using a modem. If your dialysis center uses
this option, you are given a modem to install with your HomeChoice PRO APD
System. Read the instructions that were provided with your modem for modem
operation.
Follow the steps below to install your modem.
h
NOTE:
The modem must be Class II and approved to IEC/EN, CSA/UL
60950-1 with a TNV-1 circuit (accessible circuit is limited to 60Vrms).
h
NOTE:
Maintenance of the external modem must only be performed by
qualified Baxter Service personnel. See 2.2, Numbers to Call for
Assistance, on page 2-1.
h
NOTE:
The modem must be located at least 5 feet (1.5 meters) away from the
patient.
Steps to install your modem
1. Unpack the modem.