18. Troubleshooting
18-4 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
18.2.1 Types of Alarms
There are three (3) types of alarm messages that can occur during therapy:
Auto Restart
These alarms can be corrected by reading the display screen and checking
for the problem explained in the alarm message.
You do not need to press any buttons to correct these alarms.
The alarms beep 3 times, then the system continues therapy.
If the alarm continues, the system beeps 6 times then continues
therapy.
After an Auto Restart Alarm occurs twice, the alarm becomes a Manual
Restart Alarm.
Manual Restart
These alarms must be corrected by following the instructions in
18.3, Alarms.
The alarms beep continuously until you press STOP.
Press STOP, correct the problem, then press GO to continue your
therapy.
System Error
These alarms are caused by problems inside the system. Contact Baxter
for assistance when these occur. Baxter Technical Assistance is available
24 hours a day for HomeChoice/HomeChoice PRO APD System users. Call
the number located in 2.2, Numbers to Call for Assistance, on page 2-1.