Red Cycler 5C Instrucciones de operaciones Pagina 319

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 377
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 318
18. Troubleshooting
18-64 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
If the power is restored within 30 minutes:
The system automatically resumes therapy without any alarms.
During the first 30 minutes of a power failure, press any key to retrieve
therapy information. Press
STOP and GO to start therapy after power is
restored.
If the power is not restored within 30 minutes:
An alarm occurs after 30 minutes without power. Press STOP to mute the
alarm.
Press V to retrieve therapy information.
If power is restored within approximately two hours:
The therapy can be restarted. An alarm occurs again when power is
restored.
Press STOP and GO to restart therapy.
If power is not restored within two hours:
You must end your therapy. See 18.6, End Therapy Early Procedure, on
page 18-55.
Restart your therapy from the beginning, if needed.
Vista de pagina 318
1 2 ... 314 315 316 317 318 319 320 321 322 323 324 ... 376 377

Comentarios a estos manuales

Sin comentarios