11. Operating Instructions – Prepare for Therapy
11-12 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
2. Prepare drain option.
For Drain Bag – close the clamp on the line
with the blue pull ring.
For Drain Line Extension – leave the line
clamp open.
3. If you are using a Patient Line Extension Set,
open the package and place it on a clean
surface.
4. Press
GO when you are ready to begin.
LOAD THE SET appears.
h NOTE: LOW FILL MODE IS OFF appears if your Fill Volume is less than
1000 mL and you are not using the Low Fill Mode. If this occurs, call
your dialysis center to see if your treatment should be performed in
the Low Fill Mode. Otherwise, if you are sure the Fill Volume and
therapy mode settings are correct, press
GO again to see LOAD THE
SET.
5. Push up the handle on the front of the cycler to
unlock and open the door.
h NOTE: The door must be opened within
2 minutes (30 seconds for Low Fill
Mode). If you do not open the door
within that time, press
STOP and GO
again to retract the occluder for an
additional 2 minutes (30 seconds for
Low Fill Mode).
Steps to load the disposable set (Continued)