07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 9-11
9. Operating Instructions – Change Program
Last Fill
Volume
Last Fill Volume delivered at the end of the therapy and left in
the peritoneal cavity during the day. Also called “Wet Day.”
Minimum setting is 0 or 100 mL
Maximum setting is 3000 mL
Default setting is 0 mL
Setting increments:
– 10 in the 100 to 500 mL range
– 50 in the 500 to 1000 mL range
– 100 in the 1000 to 3000 mL range
Dextrose
(default setting)
– or –
The Last Fill Volume can be the same dextrose concentration
as the other Fills or it can be different.
This setting does not appear if LAST FILL VOL = 0.
This setting only appears if you use Last Fill.
Table 9-3. CCPD/IPD Settings – Standard Fill Mode (Continued)
Setting Description