18. Troubleshooting
18-40 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
2. Press V.
The DRAIN VOLUME appears.
3. Press
V.
Initial Drain is not included in TOTAL UF.
4. Press
V.
5. Select one of the following options:
a. Press
V until MANUAL DRAIN appears.
This is the recommended option.
Press
ENTER.
– OR –
a. Press
GO to return to Drain if you do not want
to bypass.
h NOTE: When a Manual Drain ends due to low or No
Flow conditions, LOW DRAIN VOLUME
alternates with INITIAL DRAIN on the
display screen.
alternates with
– OR –
a. Press
ENTER to bypass the I-Drain Alarm.
The next FILL begins.
h NOTE: By selecting BYPASS, you indicate that you are empty. The system
considers your volume zero (0) and delivers your entire prescribed
Fill Volume.
Steps to bypass Low Drain Volume alarm
during I-Drain (Continued) Display screen