19. Technical Data
19-16 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
19.16 Determining Initial Drain Alarm Volume
Settings
Table 19-8 on page 19-17 allows you to determine the Initial Drain Alarm
(I-DRAIN ALARM) volume setting based on different percentages of the Last Fill
Volume.
To use Table 19-8, identify the row with your Last Fill Volume (mL) and read
across to the column with the desired percentage (%) to find the corresponding
Initial Drain Alarm volume setting.
EXAMPLE: If your Last Fill Volume is 2000 mL and you desire a limit that
is 85% of the Last Fill Volume, your Initial Drain Alarm volume setting is
1700 mL.
If your Last Fill Volume is between the values listed in two adjacent rows, choose
the row with lower Last Fill Volume and read across to find the corresponding
Initial Drain Alarm volume setting.
EXAMPLE: If your Last Fill Volume is 550 mL and you desire a limit that is
85% of the Last Fill Volume, your Initial Drain Alarm volume setting is
430 mL.