5. Description
5-6 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
3. DWELL
During the Dwell phase, the
system pulls fluid from the
supply bags to the
disposable cassette where
it is measured. This fluid is
then delivered to the heater
bag to replenish the
solution that was used
during the previous Fill
phase. This is done to warm
the fluid in the heater bag
to prepare for the next Fill
phase.
5.3.2 Fluid Flow During Power Failure
If there is a power failure, all valves on the cassette close so there is no fluid flow.
Valves remain closed as long as the door remains locked. See 18.9, Power Failure,
on page 18-62 for complete instructions.
5.3.3 Situations When Fluid Lines are Not Controlled
There are three (3) situations during the use of the system when the fluid lines are
not clamped by the internal occluder (behind the door of the cycler) or not closed
by the valves in the cassette.
All lines should be clamped when any of the following situations take place:
When the door is open.
During treatment setup when LOAD THE SET appears on the display
screen.
At the end of treatment when REMOVE THE SET appears on the display
screen.
During these three situations, the system does not have control of the valves and
fluid pathways. Therefore, solution can move freely between pathways. This can
result in an increased intraperitoneal volume (IIPV) situation.